Ozone Engineering
2530 Patra Drive
El Sobrante, CA 94803
(510) 758-5570
(510) 758-5571 (FAX)
cindy@ozone-engineering.com


Home Page

Product Ordering and Inquiries

Industrial Products





Pacific Ozone Tech L21 Generator

LG Series Lab Ozone Generator

Welsbach line of ozone generators from U.S. Filter/Polymetrics

Ozone Destructs


Visit Our Partners

Hydrodynesystems


API Logo

Advanced Pollution
Instrumentation, Inc.
UV Photometric Ozone Calibrator
Model 401

API Model 401 Ozone Calibrator

Features

  • Primary or transfer standard for calibration of Ozone Analyzers
  • Output concentrations from 50 ppb to 1 ppm
  • Ability to set concentration directly in ppb
  • Flow adjustable from 2 LPM to 5 LPM
  • Automatic program modes: Zero/Span; Zero/2 Span Points; Zero/5 Span Points
  • UV Lamp Feedback modes: current control; reference detector control; photometer control
  • Multitasking microprocessor controlled
  • Internal pumps
  • Powerful bi-directional RS-232 communications
  • Temperature & pressure compensation
  • Built in self checks and diagnostics
  • Non-volatile program memory
  • Digital status
  • Internal pumps for photometer and internal zero air
  • Bi-directional RS-232
  • Convenient carrying handle

Accessories

  • RS-232 cable
  • Expendables kit
  • Recommended spare parts kit
The Model 401 is a microprocessor controlled UV photometer for use as a primary or transfer standard for calibration of Ozone analyzers. The Model 401 can deliver concentrations from 50 ppb to 1 ppm at flows from 2 LPM to 5 LPM. The calibrator has two major components; the ozone generating device, and the photometer. The ozone generator is a flow controlled zero air supply passing by a UV lamp. Flow rate and lamp intensity are adjusted to deliver a specified ozone concentration by volume. The temperature- and pressure-compensated photometer (identical to the one used in the Model 400 Ozone analyzer) measures the output of the generator. Photometer output provides feedback control of the UV lamp intensity, assuring stable ozone output. A variety of automatic calibration routines can be programmed. For example, one program could generate a zero reading followed by a series of span values, while another could cycle between zero and span. Model 401 has many other versatile features, including a choice of operation from the manual keyboard, via external contact closure, remotely with RS-232, or automatically once per day on a timed basis. The diagnostics capability is the most advanced in the industry. Critical parameters are continually monitored and alarmed when out of normal. The Model 401 is versatile, simple to operate and easy to keep operational.

Voltage/frequency Options

  • 100V/50 Hz
  • 100V/60 Hz
  • 115V/60 Hz
  • 220V/50 Hz
  • 220V/60 Hz
  • 230V/50 Hz
  • 240V/50 Hz

Additional Options

  • Rack mount (19") with chassis slides
  • Rack mount only
  • Multi-drop RS-232 connection
  • Dryer cartridge for internal zero air pump

Flow Schematic

Model 401 Operation Diagram

Specifications


Ozone Generator

Flow Rate2 to 5 LPM Adjustable
Concentration Maximum1.0 ppm at 5 LPM
Concentration Minimum50 ppb at 2 LPM
Concentration Stability± 2 ppb (at 400 ppb concentration) with bench feedback

Photometer

Full Scale Rangeuser selectable 100 ppb to 10 ppm
Zero Noise.3 ppm (RMS)*
Span Noise<.5% of readinc (RMS) (Above 100 ppb)*
LDL.0006 ppb
Zero Drift<1.0 ppb/7 days
Span Drift<1.0%/day, <2%/7 days
Lag Time<10 seconds
Rise/Fall Time (95%)<20 seconds
Linearity1% of reading
Precision1.0 ppb
Bench Flow Rate800 cc/min±10%
Temperature Range5-40°C
ApprovalsCE

Physical

Height7" (178 mm)
Width17" (432 mm)
Depth27" (686 mm)
Weight60 lbs. (27 kg) with all options
Power Requirements100V 50/60 Hz, 115V 60 Hz, 220V 50/60 Hz, 230V 50 Hz, 240V/50 Hz, 250 Watts
Analog outputs10V, 5V, 1V, 100 mV

*As defined by US EPA